Conduction calorimetry: Some remarks in improved devices
Metadatos
Mostrar el registro completo del ítemAutor
Fecha
2011Materia/s
Materia/s Unesco
3305 Tecnología de la Construcción
1209.09 Análisis Multivariante
Resumen
An analysis of the systematic uncertainties in conduction calorimetry is presented. There are mainly due to the relative position of the spot of the dissipation in the sample with respect to the heat detector. In addition, wrong information is reported due to the misunderstanding between resolution and accuracy. Particular cases of miniaturized Si-based system are analyzed and some difficulties originated in conventional calorimeters from the bad interpretation of results in martensitic transformation of shape memory alloys are discussed. Copyright © 2011 Praise Worthy Prize S.r.l. - All rights reserved.
An analysis of the systematic uncertainties in conduction calorimetry is presented. There are mainly due to the relative position of the spot of the dissipation in the sample with respect to the heat detector. In addition, wrong information is reported due to the misunderstanding between resolution and accuracy. Particular cases of miniaturized Si-based system are analyzed and some difficulties originated in conventional calorimeters from the bad interpretation of results in martensitic transformation of shape memory alloys are discussed. Copyright © 2011 Praise Worthy Prize S.r.l. - All rights reserved.